Estimation of sulphanilamide by colorimetry

Aim: Estimation of sulphanilamide by colorimetry

References

  1. Indian Pharmacopoeia, 2022, Vol. I & II.
  2. Vogel’s Textbook of Quantitative Chemical Analysis, 6th Edition.
  3. Beckett & Stenlake. Practical Pharmaceutical Chemistry, 4th Edition.

Principle

Sulphanilamide, a sulfonamide antibacterial drug, forms a colored complex with diazotized reagents, which can be quantified by measuring absorbance at a specific wavelength using a colorimeter or UV-Vis spectrophotometer. The intensity of the color is proportional to the concentration of sulphanilamide in the sample.

Reagents required

  1. Sulphanilamide standard solution
  2. Sodium nitrite solution (0.1% w/v)
  3. Hydrochloric acid (1N)
  4. Ammonium sulfamate solution (0.5% w/v)
  5. N-(1-naphthyl) ethylenediamine dihydrochloride (NEDA) solution (0.1% w/v)
  6. Distilled water

Apparatus required

  1. Colorimeter or UV-Vis spectrophotometer
  2. Glass cuvettes
  3. Pipettes and burettes
  4. Volumetric flasks (10 mL, 50 mL, 100 mL)
  5. Test tubes
  6. Beakers

Procedure

1. Preparation of Standard Sulphanilamide Solution:

2. Preparation of Calibration Curve:

3. Estimation of Sulphanilamide in Sample Solution:

Calculations

From the calibration curve, determine the concentration of sulphanilamide in the sample solution:

Where:

Calculate the amount of sulphanilamide in the given formulation accordingly.

Sample result

The calibration curve was obtained by plotting absorbance vs. concentration of sulphanilamide. The equation of the standard curve was found to be:

Where:

For the given sample solution, the absorbance was found to be 0.277. Using the equation:

Thus, the concentration of sulphanilamide in the given sample was found to be 11 µg/mL.

Precautions

  1. Prepare all reagents freshly to ensure accurate results.
  2. Maintain the same reaction time for color development in all samples.
  3. Handle all chemicals, especially acids and diazotizing reagents, with care.
  4. Use the same cuvette for all absorbance measurements to reduce variability.
  5. Ensure proper calibration of the colorimeter or UV-Vis spectrophotometer before measurements.

Estimation of sulphanilamide by colorimetry